TENDERS

Annexure-I

PURCHASE OF VARIABLE PRESSURE SCANNING ELECTRON MICROSCOPE & ENERGY DISPERSIVE X-RAY ANALYSER  

Technical Specification

Resolution

2.0 nm or better in High Vacuum mode at 30kv and 4.5 nm or better in low vacuum model

Accelerating Voltage

220V to 30 KV

Probe Current

Range 1 pA to 5µA or more (continuously adjustable)

Magnification

Minimum 5x to 10,00,000x or better continuous in both high and low vacuum at all WD/scan rates

Electron Source

LaB6 filament and Tungsten hairpin filament, attachment including hardware and equivalent system.

Vacuum System

Should give ultra clean dry fast vacuum using air cooled Turbo Molecular Pump. Safety measures for electron column against any vacuum failure. Variable pressure range for chamber must be 2Pa to 600Pa minimum or more.

Stage Specification

5 axis motorized stage with movement facility for X=100mm or more, Y=100mm or more, Z=5-35mm or more, Tilt=0 to 90 0, Rotation=360 0 (continuous) with control panel.

Detectors

  1. Secondary Electron Detector for use in high & low vacuum mode.
  2. Solid state Back Scattered Electron Detector in all modes (composition, topography, shadowed) for use in high & low vacuum mode.

Safety measures in low vacuum conditions against high pressure, high moisture, contamination from samples, etc.

Image display

19 inch LCD monitor (min.) with resolution 1024X768 pixel or better.

Image store & processing

  1. Image processing resolution up to 3072X2304 pixel or more.
  2. Image analysis.
  3. Recalling achieved digital image for re-printing and re-measurement.
  4. Report generation.
  5. Image from different detectors should be viewed simultaneously in a split-up view area of the monitor.

Image format

Bmp, jpeg, tiff, etc.

Integrated computer environment operating system

With CPU minimum Intel Pentium IV with latest specification. Microsoft Windows XP/Professional or latest (licensed copy), mouse/keyboard control, facilities for scanning control, image storage, image processing, detector control, morphology, image snapper, particle/grain size analysis softwares.

  • Auto gun control and gun alignment, auto bias and auto saturation of gun condition.
  • Automatic/manual control of other features like focus, brightness and contrast, a stigmatism etc.

SEM system Disk Drives

Hard disk (160 GB or more), Floppy disk drive, CD-ROM, CD/DVD (Read & Write), optical scroll mouse & MM Key board system control.

Hard copy of image

Good quality of latest color laser printer.

Accessories required for SEM

  1. Colored Laser Printer of good quality.
  2. Column Isolation Valve.
  3. Specimen holder kits (different sizes-single stub, 6&12 stubs).
  4. Joystick box to control stage motors.
  5. CCD camera for specimen chamber inspection/infrared camera.
  6. CD Writer/Re-Writer/DVD-RW.
  7. Carbon Sputter Coater with carbon fiber head & Gold Sputter Coater.

Specification:

Control mode: Automatic and manual.

Operating pressure: 0 to 10 -2 torr or better.

 

ENERGY DISPERSIVE X-RAY ANALYSER

 

Detector

Si(Li) detector with ultra thin window and capable of detecting elements from B to U or more, Resolution-133eV or more, area-10mm 2,

  • The same EDXS detector should be compatible for high and low vacuum mode analysis
  • Safety measures in low vacuum conditions against high pressure, high moisture.

Cooling system

LN 2 with minimum detector Dewar capacity 7.5 liter or more.

Operating system

Fully PC/windows based.

PC/Hardware

P-4, 3.0 GHz or more, 1 or 2 GB RAM, 160 GB/1 TB Sata HDD, DVD Combo/48 max CD-RW, CD-ROM drive, 1.44” Floppy drive, optical sroll mouse, MM keyboard, 19 inch TFT high quality monitor, Windows XP Professional or latest (licensed copy), image processor and other useful accessories.

  • Hardware and software to control the column function and motorized specimen stage of SEM.

Software

Acquire Spectra, Advisor, Auto Column, Data Export, Elemental Identification and labeling, Element lines scan and maps, Information Management System, Navigator Mapping, point & ID, Quantitative analysis, Reporting, Spectrum Matching, Spectrum Manipulation and Display, Standardization & Image Pro Plus latest version. Automatic Gun Shot Residue Identification, Classification & Analysis Software.

Calibration standard

As per elemental range.

Accessories

  1. GSR sampling kits (five thousand)
  2. GSR analysis software capable of detection of particles with respect to size, composition, classification based on unique and typical groups of particles present, recording old BSE image and spectral data, facilities for statistical analysis of data relocation of identified particles.
  3. The operating parameters recommended for the SEM-EDXA system while using GSR analysis (as per ASTM committee E-30 adopted guidelines on GSR analysis)
  4. That the SEM while operating in the back scattered mode be capable of detecting potential of GSR particles down to a diameter of 0.5µm (attach supporting reports)
  5. Bullet & Cartridges comparison stage for common caliber.

Other related accessories

  1. UPS (compatible with the supplied SEM-EDXA) of reputed make with two hrs back up.
  2. LN2 Dewar-11 litx2

Optional Accessories

1. Peltier cool stage.

 

General Conditions:

  1. A compliance chart of the technical specifications or the quoted SEM-EDXA models with reference to our technical specifications is to be attached for each model quoted.

  2. The technical brochure/literature of the quoted models to be supplied along with the technical bid. Technical advantages of any patented parts/accessories over other conventional devices should be made available, if quoted in the technical bid.

  3. User list of the quoted SEM-EDXA models to be supplied with postal address. Email of the laboratory/contact persons.

  4. Pre-installation requirement are to be clearly mentioned (room size, environmental conditions, UPS specifications, LN2 gas cylinders for purging the specimen chamber, LN2 Dewar, power requirements, sputter coating requirements, etc.

  5. The firm will be required to analyze the supplied GSR samples using the quoted model and GSR software also by a established / Standard SEM-EDXA model and GSR software for comparison and validation of the results.

  6. The firm will be required to supply at least 5000 number of GSR sample kit including 5 pieces each of 1” stub & ½ “stub.

  7. Spares: 5 years spares like O-rings, column maintenance tools and specimen preparation kits, turbo controlled, stage rotary motor, fuse kits, R.P. oil (10 liter), LAB6 filament 2 nos, Tungsten filaments – 100 quantity.

  8. Warranty: 2 years comprehensive warranty from the date of installation is a must.

  9. Training: 3 personnel at installation site/principals factory.

  10. The company must quote 3 years comprehensive and non-comprehensive AMC rate after the warranty period.

  11. The company must quote all items for self-sustaining and trouble free installation and operation of the entire system.

  12. All operation manual, diagrams, CDs, etc must be supplied along with one set of tool box.

  13. The firm must have ISO 9001 / GMP / GLP / CE compliance.

  14. The manufacturer must certify regarding uninterrupted supply of spares and consumables for next ten years from the date of installation.

  15. All the specifications must be supported by leaflet literature and/or manufacturer’s authenticated documents.

 

Back

Website Updated on: 17 December 2007